NLF-100D
NLF-100D
NLF-100D
TECHNICAL SPECIFICATION | Value |
---|---|
Working environment | |
Ambient temperature | 5 ℃~ 40 ℃ |
Relative humidity | ≤80 % |
Atmospheric pressure | 860 hPa~1060 hPa |
Storage and transportation environment | |
Ambient temperature | -10 ℃~+40 ℃ |
Relative humidity | ≤90 % |
Atmospheric pressure | 860 hPa~1060 hPa |
Power supply | |
Voltage | 220 V±22 V~ |
Frequency | 50 HZ±1 HZ |
Input power | NLF-100 A/C 675 VA |
NLF-100 B/D/E 500 VA | |
Fuse | T10AH 250V |
Technical specification | |
Rated output pressure | NLF-100 A/B/C/D 0.35 MPa±0.05 MPa |
NLF-100 E 0.2MPa±0.05 MPa | |
Continuous output flow | NLF-100 A/C 0.35 MPa ≥45 L/min |
NLF-100 B/D 0.35 MPa ≥25 L/min | |
NLF-100 E 0.2MPa ≥25 L/min | |
Peak flow | NLF-100 A/B/C/D≥120 L/min |
NLF-100 E ≥60 L/min | |
Dew point temperature | (atmospheric pressure at 860 hPa~1060 hPa) |
NLF-100 A/B <-10 ℃ | |
NLF-100C/D/E<indoor temperature, minimum 5℃ | |
Alarm for high heat | There is a sound alarm and there is a information |
display on the screen when the compressed pump | |
temperature is over 85 ℃ ± 10 ℃ | |
Noise | ≤50 dB(A) |